Breast Implant Protocols: Technical Parameters

MRI PROTOCOL Breast Implants (silicone)

See below for saline implants**

Mayo Clinic 04/21/02 Dr. Liu
GE Signa Horizon, LX 8.3 software
These parameters may not apply to other equipment/software.

To see specifics, click here (Technical Notes
)

SERIES

#1

#2

#3

both

PRESCRIBE

LT TO RT

#4

both

PRESCRIBE

LT TO RT

#5 /#6

left/ right

#9 / #10

left / right

OPTIONAL

#7 /#8

left / right

#11 ONLY IF

DOUBLE-LUMEN

LT & RT

POSITION-PRONE

ENTRY-head 1st

COIL - breast

Chest

Scouts

FSE

IR

Water Sat

AX

FSTIR Water Sat

FSE

Water Sat

FSE

T2 FAT SAT

FSE-XL

PARAMETERS

Plane/Mode

Sequence

Options/psd

AX / 2D

SE

NP

SAG / 2D

FSE

NP,VB,FC

SAG / 2D

FSE-IR

NP,SEQ

AX / 2D

FSE-IR

NP, SEQ

AX/ 2D

FSE

NP, VB, FC

COR/OBL / 2D

FSE

VB, NP

AX / 2D

FSE

FC, NP, VB

Fse-xl

SCAN TIMING

#echo/#shots

TE / TR

TI /Flip / ETL

BW

1 echo

min full TE

400 TR

1 echo

120 TE

5800 TR

ETL 8

BW 15.63

1 echo

102 TE

4500 TR

ETL 8

TI-150

1 echo

45 TE

6000 TR

ETL 14

TI-150

1 echo

120 TE

4500 TR

ETL 8

BW 15.63

1 echo

120 TE

3500 TR

ETL 8

BW 15.63

1 echo

80TE

4500TR

ETL 12

BW 15.63

OTHER

SAT/GRX/Vasc

Gating/CINE

MultiPhase/DWI

none

WATER SAT

SAT SI

GRX

SAT Water

GRX

SAT Water

GRX

SAT SI,water

GRX

 

GRX

SAT SI, Fat

GRX

SCAN RANGE

FOV

Thickness

Spacing

Start Loc

#slices

32 fov

7 mm

skip 2.5

I55-S55

12/1

20 fov

5 mm

skip 1

 

30/1

20 fov

5 mm

skip 1

 

30/2

20 fov

5 mm

skip 1

 

26/1

20 fov

5 mm

skip 1

 

26/1

20 fov

5 mm

skip 1

 

 

20 fov

5 mm

skip 1

ACQ

Freq

Phase

NEX

Ph FOV

 

 

 

256

128

1 nex

:49

RL

H2

256

128

1 nex

:49

RL

H2

256

192

2 nex

5:12

AP

H20

Fc-FQ

Autos

256

192

2 nex

7:12

AP

FAT

 

Autos

256

192

2 nex

5:00

AP

FAT

 

Autos

256

192

2 nex

3:36

AP

FAT

Fc-SL

Autos

256

192

2 nex

2:55

RL

H20

 

AutoS

256

256

2 nex

3:30

AP

H20

FC-SL

autos

DO NOT

OBL ANY

SAG OR AX

If there is black in this scout-it has saline. Check with Rad for additional series

PRESCRIBE LEFT TO RIGHT

DO NOT OBL

 

RIGHTS SEPARATE FROM LEFTS AND LABELED

Send ALL LEFTS then ALL RIGHTS

Use Sag localizer-slice perpendicular to chest wall

DO NOT OBL

FLIP AX L-R

 

 

Angle Parallel with rib cage

Run separately

DO NOT

OBL

** IF IMPLANTS ARE SALINE: Make these changes

SAG IR- take off water sat, change CF to Water

AX FSE – Change Water Sat to FAT sat, change CF to WATER

Signal Intensities

Silicone

FAT

WATER

SAG T2 FSE

Bright

Bright

Very bright

SAG IR Water sat

Bright

Very low

Low

AX T2 FSE Water sat

Bright

Medium

Low

T2 FAT sat

Very low

Very low

Very bright